Summary
The PAMAS OLS50P is equipped with a wear resistant ceramic piston pump. The pump provides a constant flow rate of 10 or 25 ml/min at a pressure range from 0 to 6 bar.
For data transfer, the user has the choice between digital and analogue interfaces. A RS485 data interface is part of the instrument’s standard equipment for digital data transfer. The analogue 4-20 mA channel transmits data to a PLC (Programmable Logic Controller). This interface function makes the PAMAS OLS50P an effective instrument for condition monitoring.
Application
Contamination control and fluid condition monitoring
Standards and methods
ISO 11171 ISO 4402 ISO 21501-3 ISO 21501-2 ISO 4406 SAE AS4059
NAS 1638
Sample input
Pressure range: 0 to 6 bar
Constant flow rate: 10 or 25 ml/min
Volumetric particle
Calibration range:
4 to 70 µm(c) as per calibration standard ISO 11171
1 to 200 µm as per calibration standard ISO 21501
Measurement in eight size channels
Maximum particle concentration: 24.000 particles per ml at a coincidence quote of 7.8% and a flow rate of 10 or 25 ml/min
Interfaces for data transfer
Standard equipment:
- RS485
- analogue 4-20 mA interface. Parallel data transmission for four size channels or serial data transmission for all eight size channels.
Software
PAMAS POV: PAMAS Online Visualization Software for contamination control and condition monitoring